Electroacoustics - Hearing aids - Part 4: Induction-loop systems
for hearing aid purposes - System performance requirements
Abstract
IEC 60118-4:2014+A1:2017 is applicable to audio-frequency
induction-loop systems producing an alternating magnetic field at
audio frequencies and intended to provide an input signal for
hearing aids operating with an induction pick-up coil (telecoil).
Throughout this standard, it is assumed that the hearing aids used
with it conform to all relevant parts of IEC 60118. This standard
specifies requirements for the field strength in audio-frequency
induction loops for hearing aid purposes, which will give adequate
signal-to-noise ratio without overloading the hearing aid.
The standard also specifies the minimum frequency response
requirements for acceptable intelligibility. Methods for measuring
the magnetic field strength are specified, and information is
given on appropriate measuring equipment (see Annex B),
information that should be provided to the operator and users of
the system (see Annex C), and other important considerations.
This standard does not specify requirements for loop driver
amplifiers or associated microphone or audio signal sources, which
are dealt with in IEC 62489-1, or for the field strength produced
by equipment, such as telephone handsets, within the scope of
ITU-T P.370. This third edition cancels and replaces the second
edition published in 2006.
This edition constitutes a technical revision which includes the
following significant technical changes with respect to the
previous edition: Addition of Annexes G, H and I where more
information is provided about practical considerations and methods
of measurement.
This consolidated version consists of the third edition (2012) and
its amendment 1 (2017). Therefore, no need to order amendment in
addition to this publication.